SH5111 A/B/C | Fit to the relevant standard requirements of IEC60695 and IEC335 | 0.5/0.75/1 | 1200 (width) x1020 (depth) x 900(height) | 50 | 1.The temperature of the glow-wire can be adjusted from 500 to 1000℃, and the temperature tolerance for error is 500~750℃±10℃,750~1000℃±15℃,the accuracy of temperature measurement instrument is± 0.5 class. Temperature measurement using armored 0.5 to 0.55mm Ni-Cr (K) imports thermocouple. 2.Glow-Wire imposed on the samples the power of 0.8N to 1.2N, and the restrict indentation depth is at least 7 mm 3.The time the glow- Wire impose on the samples (Ta) is adjustable from 0 to 99m and 99s, Generally, according to the standard requirement set for 30 seconds, and also have the timer to measure the ignition duration time (Ti) and the flame extinguished time (Te), the time error is ± 1s . 4.The operation of the equipment is simple, the temperature measurement system is reliable and automatic temperature compensation. | 0.5 Armored Thermocouple |